Patterson group · UCSB · plan view schematic
Inelastic Recoil Spectroscopy — apparatus map
Cryogenic ion trap · single-molecule inelastic recoil spectroscopy. Click any component for how it works, the numbers, and the design reasoning.
Crystal
1 × ⁸⁸Sr⁺ + 1 molecular ion
Trap drive
Ω = 2π · 1.63 MHz · q ≈ 0.2
Trap pressure
~4×10⁻⁷ Torr cryo · 10× warm
schematic · not to scale · hover to identify, click for detail
532 load
422 / 1092
mid-IR
plan view · from above · not to scale
GM cryocooler
SRDK-415D
5.7 K
60 K
300 K
resonator ×8
5 MHz LPF
Rigol AWG
11 Vpp
loading
recoil
Ne
He
13 K
EC
EC
Paul trap
+
CCD (30%)
PMT (70%)
below trap · out of plane
Extrel QMF
leak valve
EI ionizer
Sr–Al target
532 nm
Minilite II
CEM
422 nm
MOGLabs ECDL
Rb-referenced
1092 nm
DFB repump
mid-IR OPO
~6 cm⁻¹
optics bench · 295 K
INELASTIC RECOIL SPECTROSCOPY — PLAN VIEW
PATTERSON GROUP · UCSB
REF: RSI 94, 075105 (2023)
REV 2026-07
SCALE: NTS
532 load
422 cool
1092 repump (dashed)
mid-IR (dashed)
60 K
5.7 K
13 K gas line
trap electrodes
windows